Calog Loadcell

The Calog Loadcell Calibrator is ideal for testing 4 and 6 wire loadcells, checking insulation breakdown, weighing system calibration and simulation.

It can test a loadcells' zero balance, input resistance, output resistance, bridge balance, 50V DC insulation test from bridge to shield, bridge to housing and shield to housing.

The Calog - Loadcell Calibrator is a hand-held precision calibrator specifically designed for load cell testing, including insulation breakdown, and weighing system calibration or stimulation. Simply connect a strain-guage load cell to the spring terminals and press "TEST LOADCELL". The backlit graphics will display four or six wire, zero balance, input or output resistance and bridge balance. It can then carry out a 50 VDC insulation test form bridge to shield, bridge to housing and shield to housing.

For system testing and calibration the Calog - Loadcell Calibrator can simultaneously measure mV (with mass display), excitation voltage and mA. In source mode, used for load cell amplifier pre-calibration testing, it can source mV (with mass display), measure excitation and measure mA. In simulate mode it sources a mA signal to the control room or SCADA. The display can be in mA, mA and mass or mass and mA.

Features of the Load Cell Calibrator

Measure 0 to 2000 ohms with and accuracy of 0.5 ohm and a resolution of 0.1 ohm
Measure, source and simulate 0 to 24 mA with 0.01% accuracy and 1 mA resolution
Measure 0 to 12V with 0.005% accuracy and 1mV resolution
Measure -4.5 mV to 35 mV with 0.005% accuracy and a 1 mV resolution
Continuity test function
Measure mV or measure mV and mass
Source mV or source mV and mass or source mass and mV
Simulate mA or simulate mA and mass or simulate mass and mA
Displays values in mV and mass or mass and mV
Built-in 24V power supply for mA loops
2 year guarantee (batteries not included)
Carry case, charger and test leads supplied as standard
Technical Specifications of the Loadcell Calibrator

Loadcell Test
4 or 6 wire detection and testing of 4 or 6 wire if it is functioning correctly, zero balance, input ohms, output ohms, balance + ohms, balance Calog - ohms testing.
Insulation test
Bridge to shield, bridge to housing, shield to housing @ 5000 mohm or less with an accuracy of 1%.

Analogue Input Ranges Impendance Resolution
0 to 2000 ohm 0.1 ohm
0 to 24 mA Input Impedance 100 ohm
1 microA
0 to 12 V Input Impedance 500 kohm 1mV
-4,5 to 35 mV Input Impedance 500 kohm 11 microV
Analogue Output Ranges Impendance Resolution
0 to 24 mA Output Impedance 600 ohm 1microA
-4,5 to 35 mV Output Impedance 500 ohm 1microV 

Datasheet